Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Mobile phone back cover haze testing quantifies the light-scattering behavior of smartphone rear covers made from glass, polycarbonate, PMMA, and composite materials, ensuring that surface finishes such as anti-glare etching, frosting, and optical coatings meet stringent appearance specifications. Performed in accordance with internationally recognized standards including ASTM D1003, ISO 14782, and GB/T 2410, the test delivers objective haze, transmittance, and clarity data essential for material selection, process validation, and reliability assessment. This article provides a comprehensive technical overview of haze testing principles, sample preparation, test procedures, acceptance criteria, and practical applications for mobile device rear housings.

Technical Overview

Haze is defined as the percentage of transmitted light that, in passing through a specimen, deviates from the incident beam by more than 2.5 degrees due to forward scattering. For transparent materials, haze is calculated as the ratio of diffuse transmittance (Td) to total luminous transmittance (Tt), expressed as a percentage. In the context of smartphone manufacturing, the back cover is one of the most visually prominent components, and its optical characteristics directly influence perceived quality, texture consistency, and user experience.

Modern mobile phone back covers are manufactured from a variety of substrates, including high-alumina cover glass, glass-ceramic, polycarbonate (PC), polymethyl methacrylate (PMMA), and multi-layer composite sheets. These substrates undergo secondary processing such as anti-glare (AG) chemical etching, sandblasting, texture molding, spraying, frosting, and anti-fingerprint (AF) coating. Each of these processes alters the surface microstructure and therefore changes the haze value. Haze testing provides a quantitative basis for controlling these processes and confirming that finished covers match the approved optical appearance standard.

Low haze is expected for glossy, transparent covers where optical clarity is the design intent, while higher haze values are deliberately engineered for matte, frosted, or anti-glare finishes that diffuse reflections and resist fingerprint visibility. Haze measurement is also widely used as a reliability indicator, since haze shifts after abrasion, coating wear, or environmental aging reveal degradation of the surface treatment.

  • Haze (H) = (Diffuse transmittance Td / Total transmittance Tt) x 100%
  • Primary reference standards: ASTM D1003, ISO 14782, GB/T 2410, JIS K 7361-1
  • Key parameters: haze, total luminous transmittance, diffuse transmittance, parallel (specular) transmittance
  • Typical instruments: integrating-sphere hazemeters and UV-Vis spectrophotometers

Testing Samples

Glass and Glass-Ceramic Back Covers

Cover glass samples, including 2.5D flat covers and 3D curved covers with AG-etched or frosted surfaces, are tested to verify etching uniformity, coating transparency, and batch-to-batch consistency. For 3D curved covers, measurement must be taken on flat or gently curved regions that fully cover the instrument aperture, and multiple positions should be measured to characterize uniformity across the panel.

Polymer and Composite Back Covers

Polycarbonate, PMMA, and PC/PMMA composite sheets used in molded or laminated back covers are tested to evaluate texture clarity, printing or metallizing transparency, and base-material optical purity. Specimens should be free of surface defects, streaks, and inclusions that could locally distort the haze reading.

Coated, Printed, and Finished Specimens

Samples carrying AF, AS (anti-smudge), hard-coat, or decorative coatings can be tested to confirm that the coating does not introduce unacceptable light scattering. Test specimens should be conditioned in a standard laboratory atmosphere of 23 ± 2 °C and 50 ± 10% relative humidity (per ISO 291 or GB/T 2918) for at least 40 hours prior to measurement. Both surfaces must be cleaned with lint-free wipes and ionized air to remove dust, fingerprints, and residual solvents, as surface contamination is a major source of measurement error.

  • Recommended flat specimen size: at least 50 mm x 50 mm, or a circular disc that fully covers the measurement port
  • Thickness should be reported, since haze readings are thickness-dependent for scattering materials
  • Test at least three to five positions per specimen and report the average and range

Testing Items

A comprehensive haze test program for mobile phone back covers typically includes the following measured and derived quantities:

  • Haze value (H, %) — the primary acceptance parameter for matte, frosted, and AG finishes
  • Total luminous transmittance (Tt, %) — overall light throughput under standard illuminant D65 (or illuminant A per ASTM D1003)
  • Diffuse transmittance (Td, %) — scattered component of transmitted light
  • Parallel (regular/specular) transmittance (Tp, %) — unscattered component; clarity correlation parameter
  • Haze delta (ΔHaze) — change in haze after abrasion, coating wear, or environmental aging tests, used for reliability qualification

Typical specification limits are defined by the product drawing or customer standard. Glossy transparent glass covers are commonly required to exhibit haze below 1–2%, while AG or frosted designs may target engineered haze ranges (for example, 20%–60%) together with minimum transmittance requirements to balance anti-glare performance with clarity.

Testing Methods

Hazemeter Method (Integrating Sphere)

The hazemeter method, described in ASTM D1003 (Method A) and GB/T 2410 (Method A), uses a double-beam integrating sphere optical system. The instrument first measures total transmittance with the specimen against the integrating sphere, then measures diffuse transmittance using a light trap to exclude the specular (parallel) component. Haze is computed automatically as Td/Tt x 100%. The instrument is calibrated with air as the zero reference and with certified haze and transmittance reference standards traceable to national metrology institutes.

Spectrophotometer Method

The spectrophotometer method, covered by ASTM D1003 (Method B), ISO 14782, and JIS K 7361-1, scans the transmittance spectrum across the visible range and calculates luminous quantities by weighting with the spectral distribution of standard illuminant D65 or A and the CIE photopic observer function (2° or 10°). This method allows separate evaluation of narrow-angle and wide-angle scattering, which is useful for distinguishing fine surface texture from bulk scattering.

General Test Procedure

  • Verify laboratory conditions: 23 ± 2 °C, 50 ± 10% relative humidity
  • Warm up and calibrate the instrument; verify with certified reference tiles before and after measurement
  • Clean the specimen surfaces; inspect for scratches, dust, and fingerprints
  • Place the specimen flush against the transmission port with the outer (user-facing) surface oriented consistently
  • Record haze, total transmittance, diffuse transmittance, and parallel transmittance at multiple positions
  • Report the average values, standard deviation, instrument model, standard illuminant, observer angle, and the reference standard applied

For reliability-related haze evaluation, specimens are first measured in the as-received condition, subjected to the designated stress test (for example, Taber rotary abrasion, RCA abrasion, steel-wool rub, alcohol rub, or damp-heat aging), and then re-measured under identical conditions. The reported ΔHaze value quantifies surface degradation of the coating or finish.

Manufacturers may perform haze testing in-house for routine process control and engage accredited third-party laboratories for independent verification, supplier qualification, and dispute resolution.

Applications

Haze testing of mobile phone back covers supports quality assurance and engineering decision-making throughout the product lifecycle:

  • Incoming material inspection of cover glass sheets, polymer films, and composite laminates
  • Process validation of AG etching, sandblasting, frosting, texture molding, and spraying parameters
  • Coating development and uniformity control for AF, hard-coat, and decorative optical layers
  • Reliability qualification through ΔHaze measurement after abrasion, chemical exposure, UV, and damp-heat aging tests
  • Golden-sample comparison and batch consistency monitoring during mass production
  • Failure analysis of appearance defects such as mura, blotchiness, and non-uniform frosting

FAQ

What is the difference between haze and transmittance?

Transmittance measures the total percentage of incident light that passes through the specimen, while haze measures only the scattered portion of that transmitted light. A frosted AG back cover can have high transmittance and high haze simultaneously; a clear glossy cover has high transmittance and very low haze.

Which standards govern haze testing of back covers?

The most widely applied standards are ASTM D1003, ISO 14782, GB/T 2410, and JIS K 7361-1. All define the 2.5-degree scattering criterion and specify illuminant, observer, and calibration requirements for reproducible results.

Can curved 3D glass back covers be tested?

Yes. Measurement is performed on flat or minimally curved regions that fully cover the aperture. Multiple positions along the cover are measured, and the reported haze range characterizes process uniformity across the curved surface.

What are typical acceptance criteria?

Criteria are product-specific. Glossy transparent covers commonly require haze below 1–2%, whereas engineered matte or AG finishes are specified within a defined haze window, typically paired with minimum total transmittance and ΔHaze limits after reliability stress.

How can measurement error be minimized?

Control the laboratory atmosphere, condition specimens before testing, clean both surfaces thoroughly, calibrate with certified reference standards, fix the measurement orientation, and average multiple positions. Reporting the specimen thickness and the exact standard used also improves comparability between laboratories.